State requirement record — Georgia · Last verified Jul 2026
Georgia Dental Sterilization & Spore Testing Requirements (2026)
Here's what Georgia requires for sterilizer monitoring and record retention — verified against the state's primary source — plus what an inspector checks and a free printable log sheet.
Requirement on record
Last verified Jul 2026.
At least weekly (CDC standard)
Georgia incorporates the CDC infection-control guidance for dentistry by reference, which requires biological (spore) testing of each sterilizer at least weekly.
3 years
Georgia requires sterilization / spore-test records to be retained for at least 3 years.
Ga. Comp. R. & Regs. R. 150-8-.01 (Unprofessional Conduct Defined) subsecs. (a) and (h)(5)
Informational only — not legal advice. Verify current requirements with your state dental board.
§ 01 — Inspection record
What an inspector checks in Georgia
When a Georgia board inspector or surveyor reviews a practice's sterilization records, they're confirming the monitoring actually happened and is documented. Expect them to look for:
- 01A spore-test log showing at least weekly (cdc standard) biological monitoring of every heat sterilizer in use.
- 02Records retained for at least 3 years and available on site.
- 03Chemical-indicator results recorded for processed loads, plus mechanical (time/temp/pressure) confirmation.
- 04Documented corrective action for any failed spore test, including retest and instrument recall.
- 05Compliance consistent with Ga. Comp. R. & Regs. R. 150-8-.01 (Unprofessional Conduct Defined) subsecs. (a) and (h)(5).
§ 02 — Questions of record
Georgia sterilization FAQ
- How often do dental practices in Georgia need to run a spore (biological) test?
- Georgia incorporates the CDC infection-control guidance for dentistry by reference, which requires biological (spore) testing of each sterilizer at least weekly.
- How long must Georgia dental offices keep sterilization and spore-test records?
- Georgia requires sterilization / spore-test records to be retained for at least 3 years.
- What regulation governs dental sterilization monitoring in Georgia?
- In Georgia, sterilization monitoring is governed by Ga. Comp. R. & Regs. R. 150-8-.01 (Unprofessional Conduct Defined) subsecs. (a) and (h)(5). ClaveLog has verified this against the primary source.
- What should a Georgia practice do if a spore test fails?
- Follow the CDC protocol: immediately take the affected sterilizer out of service, review the load, and re-test. Re-process and recall any implicated instruments as directed, document every step with dates and initials, and only return the sterilizer to service after a passing biological test. Keep this corrective-action record with your Georgia sterilization log — inspectors look for it.
Keep your Georgia sterilization records inspector-ready — automatically.
ClaveLog logs every autoclave load from a phone in seconds, tracks spore tests from any lab, and prints a board-ready packet in one click. Stop printing sheets — keep it digital.